PMT - Coggle Diagram
PMT
SPECIFICHE
SENSITIVITA' SPETTR CATODICA
CORR FOTOEL CHE LASCIANO FOTOCAT/POT RAD A TOT LUNG D'ONDA (A/W)
LEGATA ALLA Q.E.
CORRENTE DI BUIO
CORR ANODICA CON FOTOCAT NON ILLUMIN
SENSITIVITA' SPETTR GLOBALE
CORR ANODICA/POT RADIANTE A TOT LUNGH D'ONDA INC SU FOTOCATODO (A/W)
TEMPO DI SALITA ANODICO
T NECESS X IMPULSO DA 10% A 90% VALORE PICCO SE FOTOCAT ILLUM CON FLASH
SENSITIVITA' LUM CATODICA
CORR FOTOELETTR CHE LASCIANO FOTOC A TENS LAV/FLUSSO LUM (A/LM)
LARGHEZZA IMPULSO ANODICO
FWHM IMPULSO USCITA IN RISPOSTA A FLASH
SENSITIVITA' LUM GLOB
CORR ALL'ANODO A TENS LAV/FLUSSO LUM DA SORG IN TUNGSTENO A 2856K
VALORI MAX DI TENS E CORRENTE
RIVELATORE OTTICO A >> SENSIBILITA'
TECNOLOGIA DEL TUBO A VUOTO
MOLTIPLICAT ELETTRONICO
FOTOCATODO
OPACHI
SPESS>ESC DEPTH
DEPOSITATI SU SUPPORTO SPESSO
FOTOELETTR ESCONO DA SUP SU CUI INCIDONO FOTONI
SEMITRASPARENTI
SPESS SIMILE A ESC DEPTH
FOTOELETTR ESCONO DA SUP OPP A QUELLA SU CUI INCIDONO FOTONI
MESSI SU FINESTRA TRASP (VETRO)
SI USANO MATERIALI COME CSI,CSTE, BIALCALINI E MULTIALCALINI
MATERIALI SEMICONDUTTORI
DEVE CONVERTIRE SEGNALE LUMIN << A SEGNALE MACRO E VISIBILE
CONVERSIONE PRIMARIA
EFFETTO FOTOELETTRICO SU FOTOCATODO
EMISSIONE SECONDARIA
AMPLIFICAZ DI CARICA SU CATENA DI DINODI A POTENZ >
VALORI ENERGETICI
EN FOTOEL DETERMIN DA POTENZ DINODO
EN X ECCIT ELETTR 2-3 EV
FOTOELETTR ESCONO CON EK<1EV
ELETTR CON EK >W
PROCESSO
FOTONI PASSANO IN FIN INGRESSO
INTERAZ ATOMI FOTOCAT E EFFETTO FOTOEL
FOTOEL ACCELL E FOCUS, ESTRAGGO ELETTR SECONDARI
ELETTR SECOND RACCOLTI ALL'ANODO
FOTONE ASSORB E EN A ELETTR MATERIALE FOTOCATODO (3EV LUCE BLU)
ELETTR VA VERSO SUP E PERDE EN
PERDITE > PER AVERE MAX ESCAPE DEPTH (25NM)
ELETT ABBANDONA SUP FOTOCAT E SE EN SUFF SUPERA WORK FUNCTION (1-2EV SEMIC E 3-4EV METALLI)
1 more item...
PROCESSO DI FOTOEMISSIONE
RUMORE TERMOIONICO
< ELETTR HANNO KT>W
FALSO CONTEGGIO
PER RISOLVERE PMT IN FRIGO O AL BUIO
NELL'ANALISI DATI VEDO AMPIEZZA IMPULSI USCITA
VA RIDOTTO A CASI DOVE VEDO SEGNALI DA FOTONI SINGOLI
AUMENTA SE FOTOCATODO > (> FOTONI)
TEMPO DI TRANSITO
DIFF TEMPORALE MEDIA TRA TEMPO ARRIVO FOTONE SU FOTOCATODO E RACCOLTA ELETTRONI ANODO
VA DA 20-80 NS (VEDI GEOMETRIA PMT)
SPREAD
DURATA TEMP IMPULSI RACCOLTI A ANODO
LARGHEZZA DELLA GAUSSIANA
VALOR MEDIO E' T. MEDIO TRANSITO ELETTR
USO CALCOLO SEMICLASSICO
QUALCHE 100IO DI PICOSEC
FATTORI DA CUI DIPENDE
DISTR V INIZ FOTOELETTR (NON SEMPRE TRASCURABILE)
USO DDP > TRA FOTOCAT E 1 DINODO
N FOTOELETTR INIZ (SE >, < DISPERSIONE)
DISPERS VA COME 1RADN (> FOTOEL, < DISP)
DIFF DI CAMMINO FOTOEL (VEDI SUP DA CUI ABBANDONANO FOTOCATODO)
USO FOTOCATODI INCURVATI
SORGENTI DI NOISE
AFTERPULSES
IMPULSI SPURI RITARDATI RISPETTO AL SEGNALE
EMISSIONE LUCE ULTIMI STATI PMT
IMPULSI DA SINGOLO FOTOELETTR
RITARDO DI 20-50NS (TRANSITO ELETTR)
GAS NEL PMT (VUOTO NON PERFETTO)
IONI + DEL GAS VANNO VERSO FOTOCAT
LIBERANO 10-100 ELETTR
GRANDE RITARDO (DA 100 NS A OLTRE 1 MICROSEC
NASCONO NEGLI ULTIMI STADI (> ELETTR)