Please enable JavaScript.
Coggle requires JavaScript to display documents.
MC3 Settings - Coggle Diagram
MC3 Settings
Service
User profiles
create new/change/delete admin profile: AU
Create/delete clinical profile: AU
Vygon user management
Create new/change/delete Medwin profile: MED
Initial Vygon login fixed credentials
Initial MED login with fixed credentials
Create new/change delete Vygon profiles: VYG->MED
Create first time admin profile: everyone in presence of Vygon Sales
Reset admin profile: VYG->MED
Update software R/W: AU->VYG->MED
Title of the menu : "Update"
Last update
Update button
(SW rev, IFU, new languages)
List of software
Diagnostic and logs: AU->VYG->MED
Last alarm
List of events
Error log: TB discussed the level of details
Alarm log: TB discussed the level of details
User action log: TB discussed the level of details
Network log: TB discussed the level
Save log
User_TLOG_diag
Feature settings and status: AU->VYG->MED
Confirmation popup for settings and calibrations changes
PRAM
Pressure calibration
ABP
CVP
Analog
Pressure verification
Analog
ABP
CVP
Settings
Blood pressure excitation voltage
Status
Standard monitor actual voltage
N PRAM
Verification
Puse Oxymeter
Sphygnommanometer
Calibration
Sphygnommanometer
Settings sub menu not needed for this version
Status
sensor connection & non in status (including OEM module)
IKORUS
Settings
L parameter
IR current
HR min/max
Verification
Similar to PRAM, we connect 4 tests connector to the machine and we check that the values are correct
Status
Probe calibration value (read only)
Global factory reset including feature settings, config and User_TLOG_diag
Any future features in EXT and X1 (l1, l2 and l3 will be present as new branches here
Manufacturer: MED
Enable disable features dynamic
Reset credit
Save TLOG_RAW (download raw signal file and anonymous CLOG)
Save global TLOG_diag -> Manuf_TLOG_diag
Tests: AU->VYG->MED
L2/L3 Comms and power status
Backlight
RFID comm test
Alarm leds
Buzzer
Internal com including L1
Memory (incl RAM/ROM)
QR-code barcode
Touchscreen calibration
UART
HDMI
USB C
USB A
Colors
Autom brightness sensor
Network
Write: AU->VYG->MED
RO: all
Ethernet
MAC adress
IP/Netmask/Gateway
DHCP enable
Enable IP v6
Wifi
MAC adress
Forget network
Enable
Select network
IP/netmaks/Gateway
Enable IP v6
HL7
Enable
Client
Server
Testing
Ping
Security
TBC with 802.1x and GE manual
Private key
CSR
Import/export certificates
Configure connection
Device information
Software
IKORUS SW version
PRAM software version
ACQ SW version
OS version
General
Including machine serial number
IFU revision number
Hardware
Including Motherboard serial
Board versions
Battery
Performance
Including CPU temp
Memory (htop)
CPU (htop)
CAN bus to delete is too complicated
Disc
General setup
Time/date
Unit
Brightness
Alarm settings latching with PW: AU->VYG->MED
Languages