Please enable JavaScript.
Coggle requires JavaScript to display documents.
Shoulder FEM - Coggle Diagram
Shoulder FEM
-
-
Amplitudes
-
Contact Forces in x, y, z direction
with respect to/defined in the Load KS
applied on the Head RP which is kinematically coupled to the inner surfqace of the Konus of the head implant
-
-
BCs
-
Glenoid Bc: Encastre was applied to the Glenoid RP
Which is kinematically coupled to the back of the glenoid.
this was applied to the glenoid implant at all steps
Translation Bc: This is of type Displacement/Rotation and was used to translate the Head implant toward the glenoid in U3 which is Z or compression along mediolateral direction
to initiate the first contact.
this is not applied on the initial step and applied first at the contact generation step with value of 2 Micro meter and a ramp amplitude.
Rotation Bc: The translation Bc which was initiated in the contact generation step was then modified in the loading step. In this step no translation is applied anymore. Instead a quasi-static abduction degree is applied on the head implant to rotate around the x-axis (UR1) of Load KS
Parts
-
Coordinate system
IGES imported, Showing the position of the loading coordinate systems with respect to the implant components. Not used in the simulation. Later a Rp is defined on top of its origin to apply the loadings from HIL/MKS
-
Material
CoCr
Elastic, isotropic, E 231000, PR 0.29
UHMWPE
Elastic, Isotropic, E 1150, PR 0.36
-
Outputs
Field
-
U, Translation and Rotation
Forces/Reaction
CF, Concentrated forces and moments
RF, Reaction forces and moments
Contact
CSTRESS, CDISP, CFORCE, CNAREA, CSTATUS
-
Interactions
Contact definition
Defined in all 3 steps. created in the intial step
modified in the contact generation step. Interference Fit was modified :question: Link Title
propagated to the loading step
Interaction properties
Normal behavior
Pressure-Overclosure defined as "Hard" Contact:question:
constraint enforcement method is Default :question:
Allow separation after contact must be activated
Tangential Behavior
Friciton definiton
Isotropic, Penalty formulation :question:
Friction Coeff, 0.07
:pencil2: Sensitivity analysis on coef. :
Constraints
Glenoid
Couples the Glenoid RP to the back surface of the glenoid throu a kinematic coupling in all directions
Head
Couples the Head RP to the Inside surface of the head Konus where the load has to be applied
Loads
-
Only applied in the loading step
Defined with respect to the Load KS
Static, general, Concentrated force
-
Steps
3 Steps are used
-
Contact creation
Static General, NlGeom on. time period 1 sec is created to initiate the contact between the implants
Loading