Please enable JavaScript.
Coggle requires JavaScript to display documents.
UltraWAVE 維修 - Coggle Diagram
UltraWAVE 維修
無微波&溫度異常(T1)
HV Xformre 損壞
magnetron& magnetron temp sensor損壞
heating Xformer 損壞
3.3-3.5VAC output
HV RESISTOR& HV CAPACITOR損壞
PWS640 board or 10A fuse損壞
st20 & st6 220vac output
therrmocoupler 異常
base load 120-140 ml
漏水
check water colling tube
water colling pressure (6bar)
電磁閥or 接頭損壞
T2 溫度異常
MAIN BOARD
Temperature sensor 損壞
cooling chiller 異常
FILTER 阻塞
溫度設定錯誤
壓力異常&無法升溫至設定溫度
pressure sensor & cable損壞
漏氣
APV
可送原廠維修
Sample 問題
non-return valve
burst disk
check steel cover
check all o-ring
檢查所有接頭與insert