Please enable JavaScript.
Coggle requires JavaScript to display documents.
meX+CR service image - Coggle Diagram
meX+CR service
Calibration
Software installation
meX+ software
meX+connection
set2CRdivarioflash
Setup and calibration
Error
Data base error
License Error
Software error
Quanto VET software
spare parts list
Trouble shooting
image problem
Image has line in middle of image
Try to blow out dust by compressed air on the mirror
bundle Cleaning
check Imaging plate
Less contrast
check PMT gain
Best beam
New Calibration
No X-ray image detected even though scan had done
check module board HV
check IP in cassette
check X-ray source
Check cal parameter
strange feature on image
Check fiber bundle
check image plate
try to make new calibration
IP casette problem
stucking noise
check IP is sliding out smoothly
check LED indicator had chnaged to Green when cassette inserted
IP stucked in scanner
open top cover and
mechanical noist motor
check Alignment
Best beam
PSD calibration
Noise
check Laser optic module
check stepping motor
Scanning problem
warm up issue
Warm up message
LPSD/RPSD
value check up
check laser module and line
casette is not recognized byscanner
check RF tag on product label
check speed plate in attachd on bottom of scanner
Software problem
No connection
USB connection failuire
check device manager if driver is requried
driver update if it is requred
try to connect other USB port
check USB cable (No LAN cable!)
TEST with other USB port